| TGA Action ID |
RC-2026-RN-00361-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Dexcom G7 Sensor
Product code : STP-AT-012
Lot Numbers: 1725204004, 1725069002 |
| ARTG Number |
472165 |
| Action Type |
Recall |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
Dexcom has recently identified two lots of Dexcom G7 sensors, originally designated as scrap and intended for destruction, were stolen during the destruction process, then sold by third parties in the United States. These lots were identified for disposal and destruction due to incomplete sterilisation and deployment failure during testing. This product has been distributed through unauthorised supply channels globally.
Lot 1725069002: G7 sensors from a lot with an elevated internal testing failure rate have an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycaemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment.
Lot 1725204004: G7 sensors that are not properly sterilised have an increased risk of skin infection.
This problem only affects the listed lot numbers; no other lots are impacted. |
| Instructions for Customers/Users |
Consumers are to: - If available, check the lot number on the G7 sensor outer box. - Alternatively, search if current or unused sensor(s) are affected based on serial number by visiting www.dexcom.com/theft-check. - Immediately discontinue use and dispose of affected products. - Replace the affected sensor with an unaffected sensor or use a blood glucose meter for glucose monitoring. - Contact Dexcom Technical Support to request a replacement. Keep record of affected device(s) serial numbers for traceability and replacement purposes. - If you have symptoms of infection or inflammation, redness, swelling, or pain, at the insertion site, contact a healthcare provider.
Wholesalers are to: - Quarantine and dispose of any stock with the affected lot numbers. - Inform Australasian Medical and Scientific Ltd of affected lots in stock. |
| Action Commencement Date |
22/05/2026 |
| Responsible Entity |
|
| Contact Information |
1300 851 056 - Dexcom Technical Support |