TGA Action ID RC-2026-RN-00360-1
Type of Product Medical Device
Product Name/Description FDR Visionary X ray System

FDR-VS-B-40-CE

Multiple serial numbers

Installed in Australia between 18-Feb-2019 and 20-Nov-2024
ARTG Number 284568
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Fujifilm has identified that, under certain conditions, the contacts of the electromagnetic contactors within the X-ray high-voltage generator may weld together, allowing current to continue flowing to the power supply circuit even after the device is switched off. This may result in overheating, damage to surrounding components, and the device becoming unusable.

A protection circuit will be implemented to prevent current flow through the discharge resistor if contact welding occurs. To date, no reports of health hazards associated with this problem have been received.
Instructions for Customers/Users A Fujifilm service representative will arrange a service visit to apply the corrective action to affected units.

Until then, customers are to:
- Turn off the system at the console in accordance with the operation manual at the end of each day
- Then turn off the three-phase breaker or isolator switch connected to the X-ray generator
Action Commencement Date 14/05/2026
Responsible Entity Fujifilm Australia Pty Ltd