| Product Name/Description |
Philips Azurion and Allura Systems
Systems: Allura Xper FD10, Allura Xper FD20, ALLURA XPER FD20 BIPLANE, Allura Xper FD20 OR Table, Allura Xper FD10/10, Allura Xper FD20/15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20, Azurion 5 M12
System Codes: 722006, 722010, 722012, 722013, 722023, 722026, 722027, 722028, 722035, 722058, 722067, 722068, 722078, 722079, 722227 |
| Reason for the Action |
Philips has identified that Hard Disk Drives (HDDs) used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service.
Problems with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorised movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Philips has not received any reports of harm associated with this problem. |