TGA Action ID RC-2026-RN-00338-1
Type of Product Medical Device
Product Name/Description Ion Endoluminal System - Vision Probe Bag

Part Number: 490127

Batch Number: DM01254324
ARTG Number 471706
Unique Device Identifier (UDI) 00886874119594
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action In January 2026, Device Technologies issued an Urgent Recall (TGA Reference: RC-2026-RN-00012-1) for specific lot numbers of the Vision Probe Bag due to a sterile pouch seal integrity problem. The scope of the recall has since been expanded to include one additional batch.

Some Vision Probe Bags used with the Ion Endoluminal System may have compromised or incomplete sterile pouch seals. This problem may occur if air becomes trapped within the pouch during manufacturing and sterilisation, potentially resulting in partial seal failure and a loss of sterility.

The problem was identified through internal quality processes, and no complaints have been reported to date.
Instructions for Customers/Users Customers are to:
- Inspect the product pouch seal of all devices with the affected batch number
- Not use any product where the seal is compromised or incomplete, and dispose of affected units in accordance with their facility’s disposal procedures.
Action Commencement Date 5/05/2026
Responsible Entity Device Technologies Australia Pty Ltd