| TGA Action ID |
RC-2026-RN-00320-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Patient Side Cart (PSC) E-Brake Retainer Pin on Da Vinci System
Product Names: ASSY, PSS, SP1098
Part Number: 380601
Serial Numbers: 10633487, 10633488, 10947466 |
| ARTG Number |
97348 |
| Unique Device Identifier (UDI) |
00886874114605 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Affected devices may have received a Patient Side Cart (PSC) suspected of having a retained E-brake retention pin, which could cause uncontrolled motion and/or restricted column range of motion. It could also cause the E-brake to fail in the event of a PSC failure.
The retainer pin is intended to be used during manufacturing and removed prior to completion of assembly.
To date, there have been no reported complaints related to this problem. |
| Instructions for Customers/Users |
Customers should read and understand the contents of the customer letter. A Device Technologies representative will be in contact with customers to arrange a service of all systems affected by this notice. |
| Action Commencement Date |
29/04/2026 |
| Responsible Entity |
|
| Contact Information |
1300 338 423 - Device Technologies Australia |