Print version of this report
TGA Action ID RC-2026-RN-00311-1
Type of Product Medical Device
Product Name/Description Aural Ventilation Tubes (Grommets)

Model Number: E106

Lot Number: 12210
ARTG Number 133977
Unique Device Identifier (UDI) Refer to Device FSN Document in Supporting Information
Action Type Recall / Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action CR Kennedy & Company Pty Ltd is the supplier of Exmoor Plastics Ltd products in Australia. Exmoor recently changed its steam sterilisation supplier for PTFE grommets. It was discovered that a small number (approximately 0.1% - 0.5%) of Aural Ventilation tubes had a compromised sterile barrier integrity. Sterility assurance could not be confirmed with 100% certainty. This resulted in the potential for compromised sterile barrier integrity which may result in the product being non-sterile at the time of use.

To date, no patient harm has been reported due to this problem.
Instructions for Customers/Users Customers are to:
· Identify affected batch and quarantine immediately to prevent further use.
· Complete the Customer Response Form below and return it to rsmith@crkennedy.com.au to organise a refund/replacement.
· Do not remove any devices that are already implanted in patients.
· Patients implanted with affected devices should continue to be monitored in accordance with standard clinical practice.

This action has been closed out on 07/07/2026
Action Commencement Date 29/04/2026
Responsible Entity C R Kennedy & Co Pty Ltd
Contact Information 0419 330 118 - CR Kennedy & Co