TGA Action ID RC-2026-RN-00304-1
Type of Product Medical Device
Product Name/Description Philips MR systems with MR Elastography (MRE)

Model Name: Ingenia 3.0T
ARTG Number 230064
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Complaints have been reported regarding MR Elastography (MRE) stiffness measurements when specific image reconstruction parameters are used with Resoundant’s algorithm. The voxel size settings in the default scan protocol are too small and may result in underestimated liver stiffness values.

To date, no injuries or adverse events have been reported.
Instructions for Customers/Users Customers are to:
- Continue to use MRE as per established clinical guidelines
- Manually update the scan protocol settings as described in the customer letter

Philips will provide updates and additional information as they become available. A representative will contact customers to schedule a Field Service Engineer to implement the software upgrade (ref: 2026-PD-MR-003), with an anticipated release date of 31 March 2027.
Action Commencement Date 12/05/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team