TGA Action ID RC-2026-RN-00301-1
Type of Product Medical Device
Product Name/Description AnchorFast Guard Oral Endotracheal Tube Fastener

Product Code: 9800, 9799

All lots
ARTG Number 94910, 516468
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Hollister has received complaints that reported pressure injuries from use of the AnchorFast device.

Hollister is reminding users about the correct use of the device, including correct patient management strategies and precautionary actions to reduce the risk of pressure injuries.
Instructions for Customers/Users Customers are to follow the Instructions for Use (IFU), in particular:
· As with any fixation device, excessive pressure introduced by the device and/or the patient’s head and/or body position (e.g. prone or side-lying) may cause dermal injury, tissue ischemia, or necrosis.
· To minimise the risk of pressure injury, inspect the patient’s lips, skin and oral cavity at least every two hours or more frequently if the patient’s condition dictates (e.g. vulnerable to fluid shifts and/or oedema).
· Use caution in patients with full or swollen lips, facial swelling, dental appliances, restorative implants, and/or loose or protruding teeth.
· Discontinue use of the device if redness, skin irritation, or pressure injury occurs.

This action has been closed out on 20/08/2026
Action Commencement Date 24/04/2026
Responsible Entity Liberty Medical Pty Ltd
Contact Information 1800 279 179 - Liberty Medical Customer Service