| TGA Action ID |
RC-2026-RN-00292-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
EXENT Analyser IQ Software. An in vitro diagnostic medical device (IVD).
Used in conjunction with: – MSR801.M Immunoglobulin Isotypes (GAM) – MSQ801.M EXENT Immunoglobulin Isotypes (GAM) Control Pack |
| ARTG Number |
454106 |
| Unique Device Identifier (UDI) |
05051700020879 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
A software problem in the EXENT Analyser IQ software has been identified whereby, under rare and specific conditions, Quality Control (QC) may not be performed but is incorrectly recorded as completed.
This may result in samples being analysed and released with a green QC status despite the absence of QC. The problem only occurs in a narrow operational scenario involving depleted previously used QC vials. |
| Instructions for Customers/Users |
Customers are to: - Confirm QC results are present for each plate before releasing results. - If QC results are absent, do not release results, repeat the run with correctly loaded QC, and contact Technical Support. - During multi-plate runs using previously used QC vials, load newly opened QC vials into Reagent Runner 2 or replace the QC in Reagent Runner 1.
Thermo Fisher Scientific is working towards a software update to address this problem.
This action has been closed out on 26/10/2026 |