TGA Action ID RC-2026-RN-00290-1
Type of Product Medical Device
Product Name/Description Arrow-Clark VectorFlow Antegrade and Retrograde Chronic Haemodialysis Catheters

Multiple product codes and batches
ARTG Number 289583, 289730
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action In some cases, the sheath introducer may not split as intended due to a design related problem.

Use of a sheath introducer that does not split as intended during removal may result in immediate procedural complications, such as resistance during withdrawal, prolonged procedure time, localised pain, minor bleeding, haematoma, or superficial tissue injury. Application of excessive force during removal may lead to vessel wall injury or significant bleeding requiring medical or surgical intervention.
Instructions for Customers/Users Customers are to:
- Immediately check their inventory for product within the scope of this recall.
- Users should cease use and distribution of affected product and immediately quarantine the affected product.
- Contact Teleflex to arrange the return of these products.
- Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product.

Teleflex will advise customers when unaffected products become available.

This action has been closed out on 31/07/2026
Action Commencement Date 29/04/2026
Responsible Entity Teleflex Medical Australia Pty Ltd
Contact Information 1300 360 226 - Teleflex Medical