Product Name/Description Molift Mover 205 Floor Lifter

Models: Mover 205 and Mover 300

Product Code: M1401

Serial Number: 593304
ARTG Number 360016
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action During load testing of a Molift Mover 300, the bolt for mounting the lifting bar broke. Investigation showed defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the Molift Mover.

Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Instructions for Customers/Users Customers are to use the provided re-work kit to correct the device.