TGA Action ID RC-2026-RN-00274-1
Type of Product Medical Device
Product Name/Description Connex Spot Monitor (CSM) devices using the Welch Allyn Service Tool

Software Version 1.11 or earlier

Serial numbers ending in 3725 and above
ARTG Number 340035
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action Inconsistent leak test failures have been reported when calibration was performed using the Welch Allyn Service Tool (WAST), the software used for device upgrades, verification, and calibration activities. The leak test is one component of the overall blood pressure calibration sequence for Connex Spot Monitor (CSM) devices. This problem affects customers who purchased Connex Spot Monitors operating on PCA 409703 30 BOM A assembly A running software version 1.54, which entered production on 12 September 2025, and who also hold an active Gold License for the WAST. When calibration is attempted on devices with this configuration using WAST v1.11, the leak test step may consistently fail, resulting in an overall calibration failure.

No patient harm has been reported to date.
Instructions for Customers/Users Customers are to download and install the latest version of Welch Allyn Service Tool software (1.11.2) at https://www.hillrom.com/en/services/welch-allyn-service-tool/

This action has been closed out on 06/11/2026
Action Commencement Date 17/04/2026
Responsible Entity Welch Allyn Australia Pty Limited
Contact Information 1800 229 837 - Baxter