| Reason for the Action |
Inconsistent leak test failures have been reported when calibration was performed using the Welch Allyn Service Tool (WAST), the software used for device upgrades, verification, and calibration activities. The leak test is one component of the overall blood pressure calibration sequence for Connex Spot Monitor (CSM) devices. This problem affects customers who purchased Connex Spot Monitors operating on PCA 409703 30 BOM A assembly A running software version 1.54, which entered production on 12 September 2025, and who also hold an active Gold License for the WAST. When calibration is attempted on devices with this configuration using WAST v1.11, the leak test step may consistently fail, resulting in an overall calibration failure.