TGA Action ID RC-2026-RN-00273-1
Type of Product Medical Device
Product Name/Description B·R·A·H·M·S PlGF Plus KRYPTOR. An in vitro diagnostic medical device (IVD).

Catalog Number: CDB859.075

Lot Numbers: 59171 & 59177

Expiry Date: 23/11/2026
ARTG Number 199700
Unique Device Identifier (UDI) (01)04260157632196(17)261123(10)59171A
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Certain KRYPTOR compact PLUS analysers may underestimate Quality Control Level 1 (QC1) in the PlGF plus assay, resulting in QC values falling outside the acceptable ±20% range. When quality control procedures are followed as per the Instructions for Use, these out-of-range results should be detected before patient results are released. However, if QC is not performed correctly, the problem may go unnoticed and could lead to falsely low PlGF patient results.

No patient injuries associated with this problem have been reported to date.
Instructions for Customers/Users Customers are to:
1) Perform daily Quality Control (QC) testing before analysing patient samples.
2) If QC results fall outside ±20%, follow troubleshooting procedures or contact technical support.
3) If QC results are within range, testing can continue as normal and previously reported results remain reliable.
4) If QC was not performed as required, prior patient results should be reviewed by a clinician as appropriate.

This action has been superseded by RC-2026-RN-00415-1.

This action has been closed out on 03/06/2026
Action Commencement Date 15/04/2026
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Contact Information 1300 735 292 - Thermo Fisher