| TGA Action ID |
RC-2026-RN-00270-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
3Dimensions Mammography System and Selenia Dimensions Mammography System
All serial numbers
Multiple model descriptions and model numbers |
| ARTG Number |
166493 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The Vertical Travel Assembly (VTA) on 3Dimensions and Selenia Dimensions systems may experience internal bolt failure over time.
The VTA contains eight internal bolts that secure the C-Arm to the assembly. In 2022, Hologic received a single report of unexpected C-Arm movement during nonpatient use. Investigation of this event determined the device malfunctioned as all eight bolts had sheared, allowing the C-Arm to rotate freely.
Loose VTA bolts can experience increased mechanical stress, which may lead to bolt failure over time. Devices with loose, missing, or broken bolts may exhibit shaking, vibration, or uncommanded C-Arm motion.
Since 2022, Hologic has received a total of one hundred fourteen (114) Medical Device Report malfunction reports. |
| Instructions for Customers/Users |
Hologic is implementing an updated recurring maintenance inspection for detecting loosened, missing, or broken bolts within the VTA. Hologic Service will reach out to customers to schedule a service appointment and perform the required inspection. If any of problems are identified, the system will be evaluated for VTA bolt replacement or VTA assembly replacement, as appropriate.
It is anticipated that the initial inspection will commence in Oct 2026 subject to the availability of tooling.
In the meantime, customers may continue to use their 3Dimensions or Selenia Dimensions for its intended use.
If at any time during system operation, customers observe unexpected C-Arm movement, they should discontinue use and contact immediately Hologic Technical Support. |
| Action Commencement Date |
15/04/2026 |
| Responsible Entity |
|
| Contact Information |
1800 264 073 - Hologic Technical Support |