| TGA Action ID |
RC-2026-RN-00266-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Captura Pro Biopsy Forceps
Reference Part Number: BF-2.2S-230-20
GPN: G47690
Lot Number: W4963424 |
| ARTG Number |
176865 |
| Unique Device Identifier (UDI) |
00827002476905 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Cook Medical has confirmed that lot W4963424 of the product labelled BF-2.2S-230-20 (Biopsy forceps without spike) contains BF-2.4SL-230SP-40 devices (Biopsy forceps with spike).
The incorrect labelling could lead to misidentification of the product and as a result the tissue may not be cut as intended.
To date, Cook Medical has not received any complaints reporting harm to the patient relating to the incorrect labelling.
Other lots and products are not impacted. |
| Instructions for Customers/Users |
Customers are to: 1. Examine inventory immediately to determine if they have affected product(s). If so, quarantine any affected products that remain unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical to receive a product credit. |
| Action Commencement Date |
15/04/2026 |
| Responsible Entity |
|
| Contact Information |
1800 777 222 - Cook Medical |