TGA Action ID RC-2026-RN-00230-1
Type of Product Medical Device
Product Name/Description OXOID Agglutinating Sera, Salmonella 9-O. An in vitro diagnostic medical device (IVD).

Item: R30957301

Batch/Lots: 3719592, 3766399, 3780497, 3797948, 6107767, 6150168, 6172741, 6196409, 6206685, 6222350, 6233858, 6266130

Manufacture Dates:18.12.2023, 02.01.2024, 18.04.2024, 24.06.2024, 26.09.2024, 01.10.2024, 01.01.2025, 17.02.2025, 20.02.2025, 20.03.2025, 01.04.2025, 17.09.2025

Expiry Dates: 30.11.2026, 31.05.2027, 10.02.2028, 02.06.2028
ARTG Number 235676
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action An internal technical investigation has identified that certain lots of OXOID Agglutinating Sera, Salmonella 9-O, do not perform to Instructions for Use (IFU) criteria. Specifically, the affected lots show cross-reactivity with Salmonella 2-O within 60 seconds, which may lead to incorrect serogroup identification.
Instructions for Customers/Users Customers are to:
- Identify and quarantine any affected lots of Thermo Fisher Scientific Oxoid Agglutinating Sera, Salmonella 9-O, and discontinue use immediately.
- Dispose of any affected stock in accordance with local, state, and national requirements.
- If the product has already been used, review test results for any anomalies

This action has been closed out on 25/06/2026
Action Commencement Date 27/03/2026
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Contact Information 1300 735 292 - Customer Care Team