TGA Action ID RC-2026-RN-00229-1
Type of Product Medical Device
Product Name/Description ARTIS pheno and ARTIS icono with VE30B and VE40B Software

Product Codes: 11328100, 10849000, 11327700, 11327600
ARTG Number 273952
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action During examination, the system may show sporadically:

1. that zero dose is administered to the patient. This sporadic behavior is only shown in fluoroscopy and/or acquisitions with one (single) frame. The dose reported is "0", but dose is actually administered to the patient; OR

2. a major increase in the dose administered to the patient. The dose reported is significantly higher than the dose actually administered to the patient.

According to current data, the expected frequency of the fault is <0.5% per system.
Instructions for Customers/Users A software update will be provided for all systems with software versions VE30B and VE40B. The patch is expected to be available from Q2 2026, and completed within 6 months from software release date. Siemens will contact customers to schedule the update.
Action Commencement Date 1/04/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre