Print version of this report
TGA Action ID RC-2026-RN-00228-1
Type of Product Medical Device
Product Name/Description ARTIS pheno shipped with VE10 updated to VE30 or VE40 Software

Product Code: 10849000
ARTG Number 273952
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action An error has been identified in the 3D program parameters for ARTIS pheno systems that have been updated from VE10 to VE30 or VE40.

During 3D runs, lighter and darker patient images may be acquired which may result in less accurate 3D reconstruction.

The operation of the system is not impacted for 2D applications.
Instructions for Customers/Users This problem will be corrected in a software update.
Siemens' service organisation will contact customers to schedule the update.
Action Commencement Date 31/03/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre