| Instructions for Customers/Users |
Customers are to adhere to the instructions for use: • Do not use an IU tube beyond its 3-year life expectancy or 300 reprocessing cycles, whichever comes first. • During the treatment preparation phase, before insertion of the applicator: Visually inspect parts for indication of damage, weakening, bending, cracks, corrosion, or general deterioration. • Inspect joints and connections on play, strength or other forms of degradation.
In addition, customers are to: • Inspect the end of the IU tube to ensure that there is no gap between the tip and the fibre tube and that the tip is not detached. • During inspection, insertion and removal: avoid the use of excessive force that could damage the IU tube. • After removal: Ensure that all applicator parts have been removed from the patient and that the tip is not detached. If users find a damaged or detached IU tube tip, remove it from clinical use and contact Elekta for a replacement. |