| TGA Action ID |
RC-2026-RN-00223-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Adhese 2 DC VivaPen Kits and DC Cannulas for VivaPen
Adhese 2 DC VivaPen Kit 2ml/100 Code: 757028 Batch: Z0BJGW, Z0BK4X, Z0BLZN, Z0BM67, Z0BPTZ
DC Cannulas for VivaPen / 100 Code: 757036 Batch: Z0BNFX |
| ARTG Number |
315212 |
| Action Type |
Recall / Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Ivoclar has identified a potential safety risk in the Adhese 2 DC product, specifically the blue DC Cannulas. During additional quality testing, an unexpected variation in the amount of activator on some cannulas was observed. Depending on the individual clinical situation, this amount may result in reduced bonding of the restoration to the tooth structure and a potential loss of the restoration.
The problem relates exclusively to the blue DC cannulas. Therefore, the Adhese 2 product can be used without problem in light-curing mode using the black cannulas or with the single dose brushes in self-curing mode. |
| Instructions for Customers/Users |
Customers are to: 1. Identify the batch number on the Adhese 2 DC VivaPen and DC Cannulas for VivaPen.
2. Separate and quarantine the materials affected in practice so that they can no longer be used for dental restorations.
3(a) Contact the Territory Manager, or the Customer Service team immediately via email at orders.au@ivoclar.com or by telephone to arrange for collection of the devices affected. (or) use the Adhese 2 VivaPen in light-curing option when used in combination with only the black LC VivaPen Cannulas
3(b) Dispose of the blue DC cannulas for VivaPen in the clinical waste bin and confirm disposal using the Customer Response Form.
This action has been closed out on 31/07/2026 |
| Action Commencement Date |
26/03/2026 |
| Responsible Entity |
|
| Contact Information |
1300 486 252 - Ivoclar Customer Service |