| TGA Action ID |
RC-2026-RN-00222-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Dexcom G7 iOS App
Dexcom G7 iOS App / G7 Watch iOS App Versions: 2.9 - 2.11.2 |
| ARTG Number |
169241 |
| Action Type |
Product Correction |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
A software error was identified within the Dexcom app. While running in the background, the app may experience a transient bug that causes EGV notifications to queue up. When the user brings the app to the foreground, they may see the EGVs and trend arrows ‘replayed’ sequentially as the queue is processed. Glucose alerts will not present to the user until the trigger EGV is replayed.
It is possible for this to occur repeatedly in the background, in which case the app would perceive signal loss and begin to deploy signal loss mitigations. This has the potential to delay glucose alarms/alerts. |
| Instructions for Customers/Users |
Customers who are on the Dexcom G7 iOS / G7 Watch iOS app (V2.9-2.11.2) must update the application to the latest version available on the App Store.
This action has been closed out on 06/07/2026 |
| Action Commencement Date |
7/04/2026 |
| Responsible Entity |
|
| Contact Information |
1300 851 056 - Dexcom Technical Support |