Print version of this report
TGA Action ID RC-2026-RN-00220-1
Type of Product Medical Device
Product Name/Description Patient-Fitted Temporomandibular Joint (TMJ) Reconstruction Prosthesis

Catalogue Number: CHG020

Lot Number: 2503131014

GTIN: 07613327626575
ARTG Number 427023
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action For certain Temporomandibular Joint (TMJ) implants, there is a potential for the recommended screw length to exceed the available bone thickness where the screw hole is positioned in an area of thin bone.

This may result in dural injury, which could lead to bleeding or cerebrospinal fluid (CSF) leakage. A CSF leak increases the risk of infection, including meningitis, and may require medical or surgical intervention to prevent serious harm.

Stryker has received one surgeon-reported complaint related to this problem, in which a screw was noted to have penetrated the cranial vault. No patient harm was reported in association with this event.
Instructions for Customers/Users Surgeons are to review the Customer Letter, which includes the following information regarding risk mitigation:
- Confirm bone thickness intra-operatively
- Select an appropriate screw length based on these measurements, including use of shorter screws where required.
Action Commencement Date 30/03/2026
Responsible Entity Stryker Australia Pty Ltd
Contact Information postmarketssp@stryker.com - Stryker