| TGA Action ID |
RC-2026-RN-00220-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Patient-Fitted Temporomandibular Joint (TMJ) Reconstruction Prosthesis
Catalogue Number: CHG020
Lot Number: 2503131014
GTIN: 07613327626575 |
| ARTG Number |
427023 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
For certain Temporomandibular Joint (TMJ) implants, there is a potential for the recommended screw length to exceed the available bone thickness where the screw hole is positioned in an area of thin bone.
This may result in dural injury, which could lead to bleeding or cerebrospinal fluid (CSF) leakage. A CSF leak increases the risk of infection, including meningitis, and may require medical or surgical intervention to prevent serious harm.
Stryker has received one surgeon-reported complaint related to this problem, in which a screw was noted to have penetrated the cranial vault. No patient harm was reported in association with this event. |
| Instructions for Customers/Users |
Surgeons are to review the Customer Letter, which includes the following information regarding risk mitigation: - Confirm bone thickness intra-operatively - Select an appropriate screw length based on these measurements, including use of shorter screws where required. |
| Action Commencement Date |
30/03/2026 |
| Responsible Entity |
|
| Contact Information |
postmarketssp@stryker.com - Stryker |