TGA Action ID RC-2026-RN-00211-1
Type of Product Medical Device
Product Name/Description Atellica CH Creatine Kinase (CK_L). An in vitro diagnostic medical device (IVD).

Siemens Material Number: 11097640

Lot Numbers: 252591, 252602, 252612

Expiry Date: 01/08/2026 & 01/09/2026
ARTG Number 176331
Unique Device Identifier (UDI) 00630414006642
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Siemens Healthineers has confirmed, through investigation of customer complaints, the potential for falsely elevated patient, quality control (QC), and/or calibrator results when using Atellica CH Creatine Kinase (CK_L) affected lot numbers on the Atellica CH and Atellica CI Analyzers. This problem may be observed with both serum and plasma sample types. QC results may or may not detect the problem.

Falsely elevated results may lead to additional follow-up diagnostic laboratory testing.
Instructions for Customers/Users Customers are to:
- Immediately discontinue use of and discard the affected kit lots.
- Review the Customer Letter with a Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable.
- Review inventory of these products to determine replacement needs and to provide information to Siemens Healthineers for regulatory reporting.

This action has been closed out on 16/06/2026
Action Commencement Date 24/03/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre