| TGA Action ID |
RC-2026-RN-00203-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
PALAMIX uno mixing system
All batches with numbers below 223060 |
| ARTG Number |
157700 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Based on long-term stability testing, in units near the end of the current three-year shelf life, the mixing rod may come loose or detach from its lock during the mixing process.
The malfunction makes it impossible to mix the cement, which may cause a delay in the operating room to the procedure, while an additional pack of PALAMIX is prepared. However, the IFU states that the amount of bone cement dough required depends on the specific surgical intervention, and at least one additional set of PALAMIX should be available before commencing the operation.
No incidents or malfunctions have been reported from use. |
| Instructions for Customers/Users |
No replacement for the products is required.
Heraeus Medical Australia will reduce the shelf life of PALAMIX from three years to 17 months. From the next batch number 223060, the PALAMIX uno new shelf life will be indicated on the product labelling.
Re-labelling or a reduction of shelf life for earlier batches is not required, as all safety-related criteria continues to be met.