TGA Action ID RC-2026-RN-00201-1
Type of Product Medical Device
Product Name/Description CLARITY II ICD System

Serial Numbers: C3023G035, C3023M038, C3024C035, C3024D036, C3024E019, C3024G051, C3025D018, C3025F003, C3025F004, C3025F005, C3025F115, C3025L030
ARTG Number 426430
Action Type Product Correction
Action Level Retail
Hazard Classification Class II
Reason for the Action A potential safety-related problem may occur when operating the CLARITY II ICD system under high-fluence conditions (= 90-100 J/cm2, 1064nm) with cryogen cooling. Under these specific operating conditions, spark or loud noise (pop sound) may be observed during treatment, associated with the interaction between high-energy laser emission and cryogen cooling timing.

The problem may increase the likelihood of localised minor burn or blister on the skin.
Instructions for Customers/Users Customers are to:
Review the updated Instructions for Use (IFU), including the guidance for high-fluence operation:
• Set Cryogen Pre time and Delay time to the same value (Pre = Delay).
• Do not use configurations where Delay is set higher than Pre.
• If spark or abnormal noise is observed during treatment, immediately discontinue the procedure and assess the patient condition and the cryogen cooling timing settings before resuming treatment.

The updated IFU will become available in May 2026.
Action Commencement Date 20/03/2026