| TGA Action ID |
RC-2026-RN-00196-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Medtronic Sphere-9 Catheter
Model Number: AFR-00001 |
| ARTG Number |
468766 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Medtronic has observed an unanticipated device interaction inducing ventricular tachycardia (VT) or ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronisation Therapy-Defibrillator (CRT-D) implanted.
Through 20 February 2026, a total of 33 incidents of VT or VF induction have occurred during RF therapy delivery with the Medtronic Sphere-9 catheter in patients with a Biotronik ICD or CRT-D. The induced ventricular arrhythmias have either self-terminated or required external defibrillation, and there have been no reported deaths or serious injuries associated with this device interaction. |
| Instructions for Customers/Users |
Customers are to: - Follow the procedure recommendations listed in the Customer Letter in addition to the Instructions for Use when using a Sphere-9 catheter with a concomitant Biotronik ICD/CRT-D.
This action has been closed out on 06/07/2026 |