| TGA Action ID |
RC-2026-RN-00194-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Olympus HF (High-Frequency) Bipolar Cable
Model Number: WA00014A
All lot numbers |
| ARTG Number |
370928 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Olympus received a complaint regarding a thermal event involving the Olympus high-frequency (HF) bipolar cable during a procedure, which resulted in a minor burn injury to the user. The investigation identified that the device had been used beyond the service life specified in the product instructions for use (IFU), leading to insulation wear, corrosion, and failure of the HF cable.
A review of global complaint data over a four-year period identified 109 complaints potentially related to thermal events, with no serious injuries reported. Of these, 17 involved minor or superficial burns, or burns to personal protective equipment (PPE). |
| Instructions for Customers/Users |
Customers are reminded to: - Not use the device beyond the defined service life of 12 months, in accordance with the instructions for use. - Inspect all HF cables for mechanical defects prior to use and replace any cable where defects are identified, regardless of the first-use date or number of uses. |
| Action Commencement Date |
18/03/2026 |
| Responsible Entity |
|
| Contact Information |
1300 132 992 - Olympus Customer Operations |