Print version of this report
TGA Action ID RC-2026-RN-00186-1
Type of Product Medical Device
Product Name/Description Philips Interventional Hemodynamic System with software version 1.4.1 or lower

System codes: 722463 and 722467
ARTG Number 380951, 423852
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action There are two potential problems with the Philips Interventional Hemodynamic System:

1) Disk Space Management

The system has an automatic disk cleanup mechanism to maintain available storage space. When the disk reaches its limit, the system deletes the oldest case data. However, other files such as export files, data snapshots and generated reports are not removed. Over time, these files continue to accumulate and can eventually fill the disk. When the Hemo Host PC disk is full, new cases cannot be opened. There is the potential that case data will not be saved, and review of patient information will not be possible after closing an ongoing case.

2) Unsynchronised System Clocks

TDCO measurements and changes in patient data files may be impacted when a case is opened simultaneously in several XperIM PCs and the system clocks of the different PCs are not synchronised. If a patient case is opened on multiple PCs and the Cath Lab PC (Live Host PC) where TDCO measurements are recorded, and the clock is behind the clock of another PC, the TDCO measurement file on the Live Host PC may be overwritten by an older version from another PC. Therefore there is the potential that case data will not be saved, or the latest patient data will be replaced by an earlier file.

Philips has received no reports of patient harm related to these problems.
Instructions for Customers/Users Philips will verify the available disk space of all affected systems and delete the accumulated data when required. In addition to performing this correction, Disk space cleaning will become a standard part of the annual Philips maintenance activity.

Not all sites are impacted by the unsynchronised system clocks problem, as it only occurs when multiple Hemodynamic workstations are connected within a hospital network and are using different time-synchronisation sources. For customers with multiple connected workstations, Philips will disable the setting in the Windows operating system that allows systems to synchronise with an external time server.

In the interim, users should check the time in all Philips Interventional Hemodynamic System PCs (including the Live Host PC). If the time is not the same on each PC, they should avoid opening the same patient case on multiple PCs at the same time.

Philips expects to start implementation of these actions in March 2026.
Action Commencement Date 18/03/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team