TGA Action ID RC-2026-RN-00184-1
Type of Product Medical Device
Product Name/Description Neonatal heat therapy devices

Lifetherm 2000 heated bed, height-adjustable
Item Number: 0215000hul200

Lifetherm 2002 resuscitation unit
Item Number: 0216001hul300

Lifetherm 2003 resuscitation unit
Item Number: 0216002hul300
ARTG Number 293363
Unique Device Identifier (UDI) (01)04260171271197(11)251200(21)0215000hul99945719
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The manufacturer has identified that the display board (Part No. 0210026v01) installed in certain Lifetherm neonatal heat therapy devices may have been incorrectly assembled during production.

As a result, the device may experience a potential malfunction of the display after an undefined period of operation. In the event of damage, the component could also potentially lead to overheating and smoke development.

No incidents or device failures have been reported to date.

This market action was undertaken prior to TGA notification.
Instructions for Customers/Users Customers are to replace affected display boards with replacement stock supplied by Loewenstein Medical Technology.

This action has been closed out on 13/04/2026
Action Commencement Date 17/03/2026
Responsible Entity Lowenstein Medical Australia Pty Ltd