| TGA Action ID |
RC-2026-RN-00182-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Efficia DFM100 Defibrillator/Monitor
SW 2.00.33
Model: 866199 |
| ARTG Number |
281589 |
| Unique Device Identifier (UDI) |
N/A |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The Efficia DFM100 Defibrillator/Monitor, with software revision 2.00.33 installed, may fail to complete its boot-up sequence. This could occur either 1) during manual boot-up if the knob is turned to AED/Manual/Pacer or 2) as part of a self-test if the device is turned off and connected to any power source.
Instead of completing the boot-up sequence, the device may repeatedly display the splash screen and will not be available for clinical use. This will be indicated to the user by displaying a solid red “X” in the Ready For Use (RFU) indicator.
If the device fails to boot, defibrillation or pacing therapy may be delayed in an emergency, potentially contributing to a failure to resuscitate.
This problem was identified via customer complaints. |
| Instructions for Customers/Users |
Distributors are to: - Quarantine impacted devices and contact their local representative.
Customers are to: - Provide power to the device when it is off so automated tests can run, as part of routine maintenance. - Regularly observe the Ready for Use (RFU) Indicator to confirm the device’s readiness. - If they identify a device that shows a solid red "X" in the RFU indicator, remove the device from service and contact their local representative.
Philips will update devices that have SW 2.00.33 installed to SW 2.00.46 or higher. |
| Action Commencement Date |
12/03/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |