TGA Action ID RC-2026-RN-00176-1
Type of Product Medical Device
Product Name/Description Leksell Vantage Arc System

Left Z-Slide: Serial Numbers: 00001–00625

Right Z-Slide: Serial Numbers: 00001–00675
ARTG Number 509585
Unique Device Identifier (UDI) Slide Left: 07340048304818, Slide Right: 07340048304801, Vantage Arc: 07340048304856
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The Leksell Vantage Arc System is part of the Leksell Vantage Stereotactic System used for stereotactic neurosurgical interventions. The Leksell Vantage Arc system features Arc Slide Left and Arc Slide Right. Each of which includes a locking mechanism to set the Z-Scale coordinates. Elekta has identified cases where the Z-Scale locking mechanism may not secure properly to the on the Leksell Vantage Arc system. Investigations confirm that this problem is associated with excessive wear on the Z-Scale locking mechanism surface.

Elekta has not received any reports of patient injury related to this problem.
Instructions for Customers/Users Elekta will contact customers to arrange for the replacement of all Arc Slide Left and Arc Slide Right components once this is available.

UPDATE 25/05/2026: A correction has now released for this problem. Customers will be contacted by their local field service engineers to correct affected products.

Until their device/s have been corrected, customers are to continue to follow the advice described in the customer letter.
Action Commencement Date 11/03/2026
Responsible Entity Elekta Pty Ltd
Contact Information 02 8907 1800 - Elekta