TGA Action ID RC-2026-RN-00171-1
Type of Product Medical Device
Product Name/Description Tempus Pro Monitors and USB 2-Channel IBP Module

Part Numbers: 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 01-2017

ARTG: 317104 (Cancelled 24/06/2025)
ARTG Number 315260
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The Tempus Pro IFU uses the term “Invasive Pressure” and reference intracranial (ICP) and bladder pressure (BDR) measurements, and the device software allows pressure channels to be labelled as ICP or BDR. However, the Tempus Pro is only cleared for invasive blood pressure (IBP) monitoring, and ICP and BDR measurements have not been tested or validated for accuracy.

There have been no complaints or reports of patient harm associated with this problem.
Instructions for Customers/Users Verification and validation testing is ongoing, customers are to:
- Refer to Appendix A for guidance on how to identify whether their Tempus Pro Monitor has IBP enabled.
- Continue using the Tempus Pro Monitor, including Intracranial Pressure and Bladder Pressure measurements, as access to real-time data may support urgent clinical decision-making.
- Routinely compare ICP and BDR readings with reference values to help mitigate potential risks associated with their use.
Action Commencement Date 6/03/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team