| TGA Action ID |
RC-2026-RN-00171-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Tempus Pro Monitors and USB 2-Channel IBP Module
Part Numbers: 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 01-2017
ARTG: 317104 (Cancelled 24/06/2025) |
| ARTG Number |
315260 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The Tempus Pro IFU uses the term “Invasive Pressure” and reference intracranial (ICP) and bladder pressure (BDR) measurements, and the device software allows pressure channels to be labelled as ICP or BDR. However, the Tempus Pro is only cleared for invasive blood pressure (IBP) monitoring, and ICP and BDR measurements have not been tested or validated for accuracy.
There have been no complaints or reports of patient harm associated with this problem. |
| Instructions for Customers/Users |
Verification and validation testing is ongoing, customers are to: - Refer to Appendix A for guidance on how to identify whether their Tempus Pro Monitor has IBP enabled. - Continue using the Tempus Pro Monitor, including Intracranial Pressure and Bladder Pressure measurements, as access to real-time data may support urgent clinical decision-making. - Routinely compare ICP and BDR readings with reference values to help mitigate potential risks associated with their use. |
| Action Commencement Date |
6/03/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |