| TGA Action ID |
RC-2026-RN-00168-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Purge Cassette for Impella
Purge Cassette, Sterile Product Codes: 0043-0001, 0043-0009
Purge Cassette, Packaged Product Code: 0043-0002
Purge Cassette, 5 Pack Product Code: 0043-0003
Impella CP Smart Assist Set Product Code: 0048-0014
5.5 Accessories AU Product Code: 1000362
All unexpired Generation 1 Purge Cassette(s) |
| ARTG Number |
324256, 379190 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The purge cassette delivers rinsing fluid to the Impella catheter. The purge fluid flows from the purge cassette through the catheter to the microaxial blood pump to prevent blood from entering the motor.
Abiomed has identified an increased risk of purge leaks in Purge Cassette (Generation 1). An updated Purge Cassette (Generation 2) has been developed with a lower risk of purge leaks.
A purge leak may lead to low purge pressure and trigger a "Purge Low Pressure Alarm". If left unaddressed, low purge pressure may lead to a pump stop. |
| Instructions for Customers/Users |
Use of a purge cassette is always required when using an Impella pump.
Customers are to: - continue using the Generation 1 purge cassette until Generation 2 purge cassette stock becomes available. - ensure increased monitoring of the purge system and refer to the Instructions for Use if a “Purge Pressure Low” alarm is triggered.
Abiomed will coordinate the product removal and replacement process with customers. |
| Action Commencement Date |
6/03/2026 |
| Responsible Entity |
|
| Contact Information |
Dfletch1@its.jnj.com, OneMD-Field-Actions@its.jnj.com - Abiomed Australia |