TGA Action ID RC-2026-RN-00165-1
Type of Product Medical Device
Product Name/Description MICS3 Angled Sagittal Saw Attachment

Catalogue Number: 210490

GTIN: 07613327523874

Lot Numbers: All
ARTG Number 475546
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Stryker has identified a potential problem with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment, which may result in a foreign object in patient.
Instructions for Customers/Users Customers are to:
- Immediately inspect inventory and quarantine any affected product.
- Return the Response Form to Stryker
- Contact the local Stryker representative to coordinate return and replacement of any affected products.

This action has been finalised on 9/11/2026.
Action Commencement Date 11/03/2026
Responsible Entity Stryker Australia Pty Ltd