TGA Action ID RC-2026-RN-00150-1
Type of Product Medical Device
Product Name/Description Olympus SOLTIVE SuperPulsed Laser System

SOLTIVE Pro SuperPulsed Laser System
Model/SKU: TFL-SLS/ R5000769

SOLTIVE Premium SuperPulsed Laser System
Model/SKU: TFL-PLS/ R5000768

TFL Premium Laser Unit
Model/SKU: TFL-CPLU

TFL Standard Laser Unit
Model/SKU: TFL-CSLU

Multiple Serial Numbers
ARTG Number 344431
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Olympus has received ninety-eight (98) complaints reporting an abrupt loss of power or functionality of the SOLTIVE SuperPulsed Laser System, or an inability to power on the unit. Investigation determined that a potential defect in the 24V power supply module in some SOLTIVE Laser units may cause the system to become inoperable.

In some cases, smoke or a burning smell may occur. By design, any smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

There is no evidence of any patient injury associated with these complaints.
Instructions for Customers/Users Customers are to examine their inventory and identify any of the affected devices listed in Appendix A.

An Olympus representative will reach out to coordinate the replacement of 24V power supply module in the affected units. In the meantime, customers may continue using their SOLTIVE Laser until the repair is completed.

This action has been closed out on 14/10/2026
Action Commencement Date 27/02/2026
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations