TGA Action ID RC-2026-RN-00147-1
Type of Product Medical Device
Product Name/Description BOND Ready-To-Use Primary Antibody Synaptophysin (27G12). An in vitro diagnostic medical device (IVD)

Product Code: PA0299-U

Lot Numbers: 83059, 83607, 83992, 84289, 84733, 85058, 85998, 86360, 86779, 87306,87568, 87814, 87997
ARTG Number 178442
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action Leica Biosystems Newcastle, has identified a documentation error in the IFU for BOND Ready-To-Use Primary Antibody Synaptophysin (27G12), PA0299-U. The Instructions for Use (IFU) incorrectly lists tonsil and prostate as the recommended positive and negative control tissues, whereas the correct controls should be normal bowel and skeletal muscle. This may lead to failed control results and invalid assays requiring repeat testing. The problem is limited to the IFU, antibody performance is unaffected, and no complaints or adverse events have been reported.
Please note the Results Expected table found in the IFU has the correct positive and negative controls and had not been impacted by this error.
Instructions for Customers/Users Users may continue using all affected lots, noting the correct positive control of normal bowel and negative control of skeletal muscle.

A new copy of the IFU can be downloaded at:
https://files.leicabiosystems.com/LBS/US/All?keycode=PA0299-U&lot-range=PA0299-U+new

This action has been closed out on 25/05/2026
Action Commencement Date 27/02/2026
Responsible Entity Leica Biosystems Melbourne Pty Ltd