Print version of this report
TGA Action ID RC-2026-RN-00144-1
Type of Product Medical Device
Product Name/Description Atellica CH Enzymatic Hemoglobin A1c (A1c_E). An in vitro diagnostic medical device (IVD)

Test code: A1c_E & A1c_H

Siemens Material Number: 11097536

All lots
ARTG Number 176642
Unique Device Identifier (UDI) 00630414220505
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Siemens Healthineers has identified a problem with the Atellica CH Enzymatic Hemoglobin A1c (A1c_E) assay, where A1c results can be falsely depressed and imprecise when the Revised C-Reactive Protein (RCRP) assay is run on the same analyser. This effect increases over time and with higher RCRP testing volumes and may affect both quality control (QC) and patient results.

Siemens Healthineers is investigating the root cause and the resolution for this problem.
Instructions for Customers/Users Customers are to:
- Confirm whether RCRP and A1c_E are processed on the same analyser; if not, no action is required.
- If processed on the same analyser, monitor A1c_E QC for negative bias and increased imprecision.
Action Commencement Date 26/02/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre