| TGA Action ID |
RC-2026-RN-00143-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Aidacare Aspire Aura and Nova Lifters
Product Codes: LSS390615 & LSS390652
All serial numbers prior to AFL-PO01260162-0000 (Aura LSS390615) and ASL-PO01260162-0000 (Nova LSS390652) |
| ARTG Number |
225798, 490791 |
| Action Type |
Product Correction |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
Aidacare were made aware of a problem with the Aura and Nova patient lifters whereby the bolts that fasten leg spread actuators to the lifter frame can gradually loosen with use over time, eventually leading to bolt detachment.
The result is one or both legs decoupling from the actuators and becoming free to open and close when push/pull forces are applied to move the lifter. This may cause difficulties when moving the lifters while a patient is suspended, potentially leading to knocking patients against carers, walls or equipment. |
| Instructions for Customers/Users |
Customers are to: - Quarantine affected stock on hand. - Follow the instructions provided with the customer letter or contact Aidacare to book a service call with Aidacare Service department to apply the correction.
This action has been closed out on 24/06/2026 |
| Action Commencement Date |
4/03/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 8706 2300 - Aidacare |