TGA Action ID RC-2026-RN-00143-1
Type of Product Medical Device
Product Name/Description Aidacare Aspire Aura and Nova Lifters

Product Codes: LSS390615 & LSS390652

All serial numbers prior to AFL-PO01260162-0000 (Aura LSS390615) and ASL-PO01260162-0000 (Nova LSS390652)
ARTG Number 225798, 490791
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action Aidacare were made aware of a problem with the Aura and Nova patient lifters whereby the bolts that fasten leg spread actuators to the lifter frame can gradually loosen with use over time, eventually leading to bolt detachment.

The result is one or both legs decoupling from the actuators and becoming free to open and close when push/pull forces are applied to move the lifter. This may cause difficulties when moving the lifters while a patient is suspended, potentially leading to knocking patients against carers, walls or equipment.
Instructions for Customers/Users Customers are to:
- Quarantine affected stock on hand.
- Follow the instructions provided with the customer letter or contact Aidacare to book a service call with Aidacare Service department to apply the correction.

This action has been closed out on 24/06/2026
Action Commencement Date 4/03/2026
Responsible Entity Aidacare Pty Ltd
Contact Information (02) 8706 2300 - Aidacare