TGA Action ID RC-2026-RN-00142-1
Type of Product Medical Device
Product Name/Description Philips Azurion Systems

Product Descriptions: Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20, Azurion 5 M12, Azurion 5 M20

Model Numbers: 722067, 722068, 722078, 722079, 722223, 722224, 722226, 722227, 722228, 722233, 722234, 722236

Software versions: R1.X, R2.X, R3.X, R4.0
ARTG Number 225815, 370112
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The Reset Geometry function resets the table and stand of the Azurion system to their default position to provide all-around access to the patient. Movement continues while the button is pressed or until the default position is reached.

To prevent unintended table movement, the system provides table lock options. When a table lock is active, any table movement request from the user interface is blocked. Depending on how the Reset Geometry button is configured, pressing this button may override the table lock.

If the Reset Geometry button is accidentally pressed, unintended table movement may cause the patient to move relative to the stand-alone third-party device or stationary accessory.

Philips has also identified that, in the Instructions for Use, the Reset Geometry button is the only method described to get all-around access to the patient during a clinical emergency, while there are other ways to get such all-around access.
Instructions for Customers/Users Customers are to:
- Familiarise themselves with the additional instructions provided.
- Confirm the Reset Geometry function that best suits their clinical workflow.
- Check the current configuration of their system(s).
- Request a change to the configuration change of the Reset Geometry function, if required.

This action has been closed out on 24/06/2026
Action Commencement Date 25/02/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team