| TGA Action ID |
RC-2026-RN-00140-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Infusomat compactplus, Infusomat compactplus P & Infusomat Space P Infusion Pump
Infusomat compactplus Article Number: 8717050
Infusomat compactplus P Article Number: 8717070
Infusomat Space P infusion pump Article number: 8713070 |
| ARTG Number |
116814 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The Infusomat is equipped with a pressure sensor designed to detect upstream occlusions. During active infusion, the system continuously monitors pressure levels and triggers a high-priority alarm to alert the user if a significant drop is detected. This mechanism ensures early detection of potential problems and supports timely intervention.
In very rare cases, if the alarm is triggered multiple times and the user mutes it each time without resolving the actual cause of the blockage (incorrectly both confirming “OK” to the “Check Upstream” display prompt and confirming “YES” to the “Is the alarm cause resolved?” display prompt upon product reset) the system may begin to treat the current (already low) pressure level as the new reference point for normal operation.
The potential health risks associated with this problem primarily relate to undetected under-infusion. No patient harm has been reported to date due to this problem. |
| Instructions for Customers/Users |
- B Braun has implemented updates to the Instructions for Use (IFU) of all affected products. - Customers should revise the updated IFUs and complete the Customer Response Form. |
| Action Commencement Date |
25/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 705 - B Braun Customer Service |