TGA Action ID RC-2026-RN-00138-1
Type of Product Medical Device
Product Name/Description Alaris neXus and Alaris Plus syringe pumps

Catalogue Numbers: 8002TIG03, 8002TIG03 -G, 8003TIG03, 8003TIG03 -G, 8005TIG03, CCneXus1, CCneXus1 -S, PKneXus1

All Serial Numbers
ARTG Number 177501
Unique Device Identifier (UDI) 10885403485268
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action BD has identified through an internal investigation that the BD Alaris neXus and Alaris Plus syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion. This is an internal component and not readily visible to the Clinical User. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9.

To date, there have been no reported serious adverse events as a result of this problem.
Instructions for Customers/Users - BD has updated the manufacturing process to specify the shipping configuration of the BD Alaris neXus and Alaris Plus Syringe Pumps.
- BD has updated the product Directions for Use which can be accessed on the eIFU website (eIFU.bd.com).
- The Technical Service Manuals have also been updated and can be accessed by qualified service personnel through the BD online service portal.

Update 18/03/26: BD has updated the Customer Letter to clarify the section on 'Actions for Biomedical Engineers'. Refer to updated Technical Service Manuals BDTM00003 Issue 4 and BDTM00010 Issue 7 for more information.
Action Commencement Date 24/02/2026
Responsible Entity Becton Dickinson Pty Ltd
Contact Information ANZ_Quality@bd.com - Becton Dickinson