| TGA Action ID |
RC-2026-RN-00138-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Alaris neXus and Alaris Plus syringe pumps
Catalogue Numbers: 8002TIG03, 8002TIG03 -G, 8003TIG03, 8003TIG03 -G, 8005TIG03, CCneXus1, CCneXus1 -S, PKneXus1
All Serial Numbers |
| ARTG Number |
177501 |
| Unique Device Identifier (UDI) |
10885403485268 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
BD has identified through an internal investigation that the BD Alaris neXus and Alaris Plus syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion. This is an internal component and not readily visible to the Clinical User. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9.
To date, there have been no reported serious adverse events as a result of this problem. |
| Instructions for Customers/Users |
- BD has updated the manufacturing process to specify the shipping configuration of the BD Alaris neXus and Alaris Plus Syringe Pumps. - BD has updated the product Directions for Use which can be accessed on the eIFU website (eIFU.bd.com). - The Technical Service Manuals have also been updated and can be accessed by qualified service personnel through the BD online service portal.
Update 18/03/26: BD has updated the Customer Letter to clarify the section on 'Actions for Biomedical Engineers'. Refer to updated Technical Service Manuals BDTM00003 Issue 4 and BDTM00010 Issue 7 for more information. |
| Action Commencement Date |
24/02/2026 |
| Responsible Entity |
|
| Contact Information |
ANZ_Quality@bd.com - Becton Dickinson |