TGA Action ID RC-2026-RN-00136-1
Type of Product Medical Device
Product Name/Description Instinct Plus Endoscopic Clipping Device

Part Number: INSC-P-7-230-S

Order Number: G58010

Lot Numbers: all INSC-P-7-230-S devices manufactured between 09 Feb 2023 through to 20 Apr 2025
ARTG Number 96698
Unique Device Identifier (UDI) 00827002580107
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Cook Medical has received reports of a malfunction that occurs when the user attempts to open the clip jaws by actuating the handle. The malfunction that may occur is the clip housing detaching from the catheter attachment and the clip remaining attached to the drive wire. The clip will extend from the catheter, but it will remain attached to the internal drive wire instead of opening the clip jaws. The clip cannot be opened if this occurs.

98% of customer reports result in low impact. The remaining 2% have resulted in minor harm requiring medical intervention.
Instructions for Customers/Users Customers are to:
1. Examine inventory immediately and cease all further distribution and use of the affected products.
2. Return the affected product(s) to Cook Medical with a signed copy of the Customer Response Form.
and email the completed form to ausrecalls@cookmedical.com.
Action Commencement Date 23/02/2026
Responsible Entity William A Cook Australia Pty Ltd
Contact Information 1800 777 222 - Cook Medical