| TGA Action ID |
RC-2026-RN-00125-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
MobileDiagnost wDR system
Model Numbers: 712001, 712002, 712005, 712006, 9890-010-89522
Manufactured before 08/2022 |
| ARTG Number |
187112 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
If a user, while cleaning or disinfecting the MobileDiagnost wDR system, uses excessive liquid or if liquid is sprayed directly on the system, it is possible for liquid to enter the edges of the handgrip buttons and cause metal corrosion or an electrical short circuit to the overlay connector (Refer Figure 1). The system has four (4) buttons on the handgrips which control the motion of each driving wheel (forwards / backwards) and if the system experiences an electrical short circuit due to this problem, the system may move on its own in slow (20 cm/s) motion. At this speed, a person or bystander can easily move away from the system without any injury. The user can easily access the Emergency Stop button and stop the system movement. The problem does not occur once the liquid dries. Philips has received 39 complaints regarding liquid ingress causing a short circuit of the fine positioning control button. Philips investigation concluded that this problem is related to the excessive use of disinfectant cleaning solutions. |
| Instructions for Customers/Users |
Customers may continue to use their system in accordance with the intended use. • If unintended motion occurs: - Press Emergency Stop button to prevent further movement - Allow unit to dry. Once dry, the system will move as intended. • During Cleaning: - Do not spray disinfectants directly on the buttons, since this may cause electrical short circuits or metal corrosion. - Follow the cleaning instructions provided in the Maintenance, Cleaning and Disposal section of the instructions for use (IFU).
A Philips representative will contact customers to schedule a time for a Field Service Engineer (FSE) to visit each site to implement the following (reference FCO71200255): • Install a plastic overlay on the grip to prevent liquid from entering the gaps around the fine positioning control buttons. • Install a firmware upgrade that stops system movement if triggered by a short circuit or extended activation of the control button.
Philips expects to release FCO71200255 by March 2026. |
| Action Commencement Date |
17/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |