Print version of this report
TGA Action ID RC-2026-RN-00124-1
Type of Product Medical Device
Product Name/Description Anaesthesia Circuits Limbo

ANES CIRCUIT, LIMBO, 108 IN, 2L BAG, AND ADAPTOR, 22/15MM OD
Code: AFPXOXXX

ANES CIRCUIT, LIMBO, 72 IN, 2L BAG, AND ADAPTOR, 22/15MM OD
Code: AFNXOXXX

BAG ADAPTER 22 MM X 22 MM X 15 MM
Code: 5156EU
ARTG Number 234578, 508065, 508365
Action Type Product Correction / Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Airlife Australia has identified that anaesthesia breathing circuits AFNXOXXX and AFPXOXXX were incorrectly packaged and shipped with SKU 5156EU (adaptor). The circuit IFU was included in the box; however, the regulated and approved IFU specific to SKU 5156EU was not included. Instead, an unregulated “Important Notice” document was placed inside the circuit case with the adaptor.

This represents labelling and documentation control nonconformance related to the 5156EU component.

This problem may lead to a potential use error in which hypoxia could occur due to incorrect therapeutic gas delivery. The absence of clear labelling and instructions may increase the likelihood of user confusion, potentially resulting in improper circuit configuration and reduced oxygen delivery.
Instructions for Customers/Users Customers are to:
- Inspect inventory immediately for the affected stock of AFPXOXXX and AFNXOXXX on hand.
- Discard the “Important Notice” document
- Include a printout of the 5156EU IFU (supplied with the Customer Letter) in the package for the adaptors.
- Instruct customers and any impacted surgeons, clinicians, or other areas that may have received the affected products to follow the IFU (supplied with the Customer Letter) if an adapter is required for connection.
- If any problems are experienced relating to this notice, retain the sample and notify Airlife immediately.

This action has been closed out on 18/08/2026
Action Commencement Date 26/02/2026
Responsible Entity Airlife Australia Holdings Pty Ltd
Contact Information 1800 519 445 - Airlife Australia