Print version of this report
TGA Action ID RC-2026-RN-00123-1
Type of Product Medical Device
Product Name/Description Imactis CT-Navigation system

HIBCC / UDI [DI] : +B681J02000
ARTG Number 445211
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of a potential problem on Imactis CT-Navigation systems that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position. An inaccurately simulated needle trajectory, if not recognised, could result in advancement of the needle to an unintended location and potential patient harm.

There have been no injuries reported as a result of this problem.
Instructions for Customers/Users Customers can continue to use the system if they do not use needles with metal handles or metal hubs.

A User Manual addendum is included in the Customer Letter that includes specific instructions to avoid the use of needles with metal handles or metal hubs.

This action has been closed out on 30/07/2026
Action Commencement Date 19/02/2026
Responsible Entity GE Healthcare Australia Pty Ltd
Contact Information 1800 659 465 - GE Healthcare