| TGA Action ID |
RC-2026-RN-00123-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Imactis CT-Navigation system
HIBCC / UDI [DI] : +B681J02000 |
| ARTG Number |
445211 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
GE HealthCare has become aware of a potential problem on Imactis CT-Navigation systems that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position. An inaccurately simulated needle trajectory, if not recognised, could result in advancement of the needle to an unintended location and potential patient harm.
There have been no injuries reported as a result of this problem. |
| Instructions for Customers/Users |
Customers can continue to use the system if they do not use needles with metal handles or metal hubs.
A User Manual addendum is included in the Customer Letter that includes specific instructions to avoid the use of needles with metal handles or metal hubs.
This action has been closed out on 30/07/2026 |
| Action Commencement Date |
19/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 659 465 - GE Healthcare |