TGA Action ID RC-2026-RN-00119-1
Type of Product Medical Device
Product Name/Description 16F Dual-Valved Splittable Sheath Introducer

Part Numbers: H965103028031/EU, H965103028041/EU, H965103028051/EU

Lot Numbers: I3270828, I3302439, I3302440

Expiry Dates: 31/03/2028, 30/04/2028
ARTG Number 283676
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Merit Medical Systems, Inc. is conducting a recall of the Merit 16F Dual-Valved Splittable Sheath Introducer due to a design defect. Specifically, the sheath introducer may not split as intended.

Merit has received customer complaints regarding this problem, some of which involve patient injury. Failure of the sheath introducer to split as intended may result in haemorrhage, foreign bodies, and procedure delay.
Instructions for Customers/Users Customers are to immediately discontinue use of the affected product and destroy it at point of use.

The other products included with the sheath introducer may continue to be used.
Action Commencement Date 4/03/2026
Responsible Entity Merit Medical Australia Pty Ltd