TGA Action ID RC-2026-RN-00118-1
Type of Product Medical Device
Product Name/Description Digital Radiography System Collimator Type R-20J

Serial Number: All
ARTG Number 314584
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action It was identified that the publication year shown on the laser classification label for the line marker option of the Collimator Type R-20J was incorrect. The label referenced the standard as published in 2001; however, the correct publication year is 2007 for IEC 60825-1. Consequently, the product did not meet the requirements of the applicable harmonised (recognised) standard.

The affected units have been distributed to hospitals since 2015.
Instructions for Customers/Users Customers are to continue using the device. Shimadzu will replace the label in the coming weeks.
Action Commencement Date 13/02/2026
Responsible Entity Shimadzu Oceania Pty Ltd