| TGA Action ID |
RC-2026-RN-00108-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Trilogy Evo, Trilogy Evo O2, and Trilogy EV300
Identifiers: IN2110X15B, IN2100X15B, IN2200X15B, AU2110X15B
ARTG: 336800 (Cancelled 07/03/2025) |
| ARTG Number |
332200 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Analysis of the impact of nebulisers on performance of Trilogy Evo Platform ventilators revealed that using non-pneumatic nebulisers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Non-pneumatic nebulisers change the airflow characteristics by adding liquid droplets, causing the ventilator’s leak estimation calculation to be incorrect. This leads to under-delivery of therapy to the patient even though the graphic user interface does not display the change. This condition may occur with or without aerosol accumulation on the flow sensor. It is also not consistent with the requirements of ISO 80601-2-121.
There have been three (3) complaints reported for this problem, including one (1) with minor injury and two (2) with no injury reported. |
| Instructions for Customers/Users |
ORIGINAL ADVICE 9/02/2026 - Customers should stop all use of non-pneumatic nebulisers, review the latest version of the User Manual Addendum, and ensure the addendum is included with every Trilogy Evo Platform ventilator in their fleet.
UPDATED ADVICE 12/05/2026 - Based on additional internal review and clinical evaluation, Philips Respironics has evaluated the risk vs benefit of continued use of non-pneumatic nebulisers (e.g. vibrating mesh nebulisers) and is revising their guidance to allow for the use of non-pneumatic nebulisers, as long as the updated instructions for correct placement are followed, and customers understand the following warning.
Specifically, similar to pneumatic nebulisers, non-pneumatic nebulisers may be used AS LONG AS they are placed in accordance with the guidance and proper location on the various types of patient circuits. See Figure 1 in the updated customer letter.
WARNING: Ventilator accuracy and performance can be adversely affected by changes to the inspired gas caused by the use of both pneumatic and non-pneumatic nebulisers, thereby posing a potential patient safety risk.
Philips will issue an updated User Manual and software which will clarify that the guidance applies to both pneumatic and non-pneumatic nebulisers. |
| Action Commencement Date |
9/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 830 517 - Philips Customer Support |