Print version of this report
TGA Action ID RC-2026-RN-00106-1
Type of Product Medical Device
Product Name/Description VITROS Immunodiagnostic Products Progesterone 2 Calibrators & VITROS Immunodiagnostic Products Progesterone 2 Reagent Pack. An in vitro diagnostic medical device (IVD).

Product Codes: 619 9455 & 619 9454

Lot Numbers: 0050, 0060, 0070, 0080, 0090, 0140, 0150, 0160

Expiry Dates: 03-Feb-2026, 04-Mar-2026, 11-Mar-2026, 31-Mar-2026, 06-May-2026, 17-Sep-2026, 28-Oct-2026, 10-Nov-2026
ARTG Number 176641
Unique Device Identifier (UDI) 10758750035373, 10758750035366
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action QuidelOrtho, through customer complaints, has identified a problem with their VITROS Prog2 Calibrators. There is an unexpected increase in signal over time in VITROS Prog2 Calibrators, Levels 2 and 3. This may cause:

o Calibration failures when attempting to calibrate impacted lots of VITROS Prog2 Reagent Pack.

o Quality control (QC) failures with results above the limit, due to a sub-optimal passing calibration, when performing QC on impacted lots of VITROS Prog2 Reagent Pack.

To date, QuidelOrtho has received 125 complaints related to this problem, with no reports of adverse effects.
Instructions for Customers/Users Customers are to:
- Follow the steps outlined in the Customer Letter to recalibrate the systems.
- Complete the Customer Response Form and return it to mikaela.rasile@quidelortho.com.
Action Commencement Date 6/02/2026
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Contact Information 1800 032 359 - QuidelOrtho Technical Services