| TGA Action ID |
RC-2026-RN-00104-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Bipolar Wristed Instruments for the Hugo robotic-assisted surgery (RAS) system
Bipolar Fenestrated Graspers CFN: MRASI0004 GTIN: 10884521739970, 10884521826700, 10884521844582
Bipolar Maryland Forceps CFN: MRASI0005 GTIN: 10884521739987, 10884521826717, 10884521844599 |
| ARTG Number |
377770 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There were complaints involving breakage of the bipolar pulley and plastic spacer components in the Bipolar Fenestrated Graspers and Bipolar Maryland Forceps used with the Hugo RAS system. Investigations show that loss of epoxy bonding, instrument interaction leading to jaw misalignment, and subsequent increased loads can result in pulley separation and failure.
These failures, associated with instruments manufactured between November 2022 and November 2025, have caused incorrect jaw actuation, uncontrolled motion, and detachment of small white plastic fragments into the patient cavity. |
| Instructions for Customers/Users |
Customers are to: - Inspect instruments prior to use and discontinue use immediately if damage, jaw misalignment, or abnormal motion is observed. - If a white plastic component is dislodged or suspected to be dislodged, it must be secured and removed from the patient cavity before proceeding; the component is not radio-opaque and radiographic imaging should not be used to locate it. - In the event of instrument failure, remove the instrument using the Disabled Instrument Workflow as outlined in the Hugo RAS User Guide and/ or Quick Reference Guide. |
| Action Commencement Date |
6/02/2026 |
| Responsible Entity |
|
| Contact Information |
0487 880 141 - Dean Kanakis, Medtronic Australasia |